Control Processing Temperature to Improve Your Mix
The LabRAM/PharmaRAM Jacketed Vessel system assembly consists of a fixed jacket, removable mixing vessel liner and lid, tubing and optional chiller / heaters. Jacketed Vessel systems bolt into place, replacing the vessel holder. One RTD temperature probe port and two vacuum line ports allow integration with other accessory systems. The system material is 316 stainless steel.
Key Processing Capabilities
- Add temperature control to mixing and processing capabilities of RAM for
- Solids/Powders – from nano-scale and higher
- Liquids/Gases
- Viscous/Slurry materials
- Coating, Milling, Grinding Processes
- Monitor target temperature of mixing material for improved results
- Automate fluid circulation control to enable hands-free operation
- Apply vacuum and temperature measurements simultaneously during mixing using on board operating controls
- Eliminate vessel changes with multiple, same vessel capabilities of LabRAM II and PharmaRAM II
Jacketed Vessel Features
- Temperature control from 302°F (150°C) heating to 0°F (-18°C) by customer supplied plant water or optional chiller / heater
- Easy installation and operation with LabRAM II and PharmaRAM II pre-installed plumbing and connections
- Equipped with RTD for temperature monitoring through mixer’s HMI
- Removable and replaceable mixing vessel liner in various sizes
- Equipped with two vacuum ports for optional vacuum systems
- See LabRAM II H Vacuum for hazardous and energetic applications
Compatible RAM Mixers
Specifications
Process Requirements | QTY | Alt. units | |
Vessel | Product vessel pressure | 3.5 bar | 50 psig |
Product temperature measurement | 1 port | ||
Product vacuum/pressure relief | 2 ports | ||
Jacket | Jacket cooling/heating | 34°- 302° F | 1° – 150°C |
Jacket pressure | 3.5 bar | 50 psig | |
User Supplied Requirements | |||
System | Heating System | Steam/Hot Fluid | |
Cooling System | Chilled Fluid | ||
Process Instrumentation and Control | Gauges, valves, heat/cold header, steam trap |
Physical Specification | QTY | |
Vessel | Volume (product + headspace) 500 ml, 16 oz. | 1 |
Face seal: lid to vessel and vessel to jacket. Viton | 2 | |
Lid | ¼” Swagelok compression fittings for RTD and vacuum. 316 SS, RTD, PTFE ferrule, cable included. | 1 |
¼” instant tube fitting for hose routing to off-board vacuum. Valve included. | 2 | |
Jacket | 1/8” NPT female half couple for heat transfer media In – Out – Drain | 2 |
Assembly | Lid to jacket mount: Sanitary clamp | 1 |
Braided SS hose to include service loop, connect outer ring to vessel. Inlet, outlet, pressure relief | 3 | |
Jacket to LabRAM mount: Dog eared flange with 1/4-20 cap screws | 4 |